浩瀚色譜采用《空氣和廢氣監(jiān)測分析方法(第四版)》P585~P587總烴和非甲烷總烴測定方法一(B)利用六通閥定量環(huán)進樣來測定環(huán)境空氣以及廢氣中的非甲烷總烴濃度,對儀器條件進行優(yōu)化,方法的檢出限為0.13 mg/m3,標樣總烴相對標準偏差為0.4%,甲烷的相對標準偏差為1.2%,總烴的加標回收率為100.9%~101.6%,甲烷的加標回收率為102.4%~102.6%,該方法操作簡單,分析速度快,實驗結(jié)果滿意。
?
名稱:MS5A甲烷柱
規(guī)格:1m*1/8,60-80Mesh
*高使用溫度:300°C
編號:20240615002
應(yīng)用:天瑞VOC,在線,便攜式
實例;固定污染源廢氣/環(huán)境空氣中甲烷的測定? 氣相色譜法
特點:氧與甲烷分離度大
?
?
Haohan Chromatography adopts the "Air and Waste Gas Monitoring and Analysis Methods (Fourth Edition)" P585~P587 Total Hydrocarbon and Non Methane Total Hydrocarbon Determination Method 1 (B), which uses a six way valve to quantitatively inject samples to determine the concentration of non methane total hydrocarbons in ambient air and waste gas. The instrument conditions are optimized, and the detection limit of the method is 0.13 mg/m3. The relative standard deviation of total hydrocarbons in the standard sample is 0.4%, the relative standard deviation of methane is 1.2%, the recovery rate of total hydrocarbons is 100.9%~101.6%, and the recovery rate of methane is 102.4%~102.6%. This method is easy to operate, fast in analysis, and satisfactory in experimental results.
Name: MS5A Methane Column
Specification: 1m * 1/8, 60-80Mesh
*High operating temperature: 300 ° C
Number: 20240615002
Application: Tianrui VOC, online, portable
Example; Determination of methane in fixed pollution source exhaust gas/ambient air by gas chromatography
Features: High separation of oxygen and methane